The Fluke 718Ex Intrinsically Safe Pressure Calibrator is a powerful intrinsically safe, self contained pressure calibrator for usage in explosion endangered areas. You can the rent Fluke 718Ex Intrinsically Safe Pressure Calibrator in Bangladesh, Reliable solution for pressure calibration in hazardous areas with accurate & safe performance
| Pressure Measurement Accuracy | ||
| Fluke 718Ex 30G | Range | -830 mbar to 2 bar (internal sensor) Over pressure: 3x FS |
| Resolution | 0.0001 bar | |
| Accuracy | 0.05% FS | |
| Functions | Zero, Min, Max, Hold, Damp | |
| Fluke 718Ex 100G | Range | -830 mbar to 2 bar (internal sensor) Over pressure: 2x FS |
| Resolution | 0.001 bar | |
| Accuracy | 0.05% FS | |
| Functions | Zero, Min, Max, Hold, Damp | |
| Fluke 718Ex 300G | Range | -830 mbar to 20 bar (internal sensor) Over pressure: 375 psi, 25 bar |
| Resolution | 0.0001 bar | |
| Accuracy | 0.05% FS | |
| Functions | Zero, Min, Max, Hold, Damp | |
| Fluke 718Ex series | Range | Via pressure module connector 29 pressure modules, 2.5 mbar to 700 bar Over pressure per pressure module specs |
| Resolution | Per pressure module specs | |
| Accuracy | To 0.025% of full span, per pressure module specs | |
| Functions | Zero, Min, Max, Hold, Damp | |
| Source pressure, built in pump range | -820 mbar to full scale | |
| For use with non-corrosive gasses. Supported Pressure Units psi, inH2O (4°C), inH2O (20°C), kPa, cmH2O (4°C), cmH2O (20°C), bar, mbar, kg/cm², mmHg, inHg Media compatibility per pressure module specs |
||
| Technical Data | ||
| Measure accuracy | Range | 0-24 mA |
| Resolution | 0.001 mA | |
| Accuracy | 0.02% rdg ±2 counts | |
| Display | LCD, 5 digit pressure and current simultaneous | |
| Environmental Specifications | ||
| Operating temperature | -10°C to 55°C | |
| Storage temperature | -40°C to 60°C | |
| Humidity (without condensation) | 95% (10°C to 30°C) | |
| 75% (30°C to 40°C) | ||
| 45% (40°C to 50°C) | ||
| 35% (50°C to 55°C) | ||
| Operating altitude | 3,000 m max | |
| Safety Specifications | ||
| Safety | ATEX II 1G Ex ia IIC T4 compliant | |
| CSA Certification I.S. Class I | ||
| Div 1 Groups A-D , T4 | ||
| Mechanical and General Specifications | ||
| Size (L x W x H) | 216 x 94 x 66 mm (with holster) | |
| Weight | Approx. 992 g (with holster) | |
| Warranty | Three years, one year for pump | |
| Shock | 1 meter drop test | |
| Vibration | Random, 2 G, 5-500 Hz | |
| Battery type | One 9-volt alkaline battery ANSI/NEDA 1604A or IEC 6LR619V | |
| Battery life | 4 to 20 hours, depending on functions used | |